Product Details | BLOOD GAS ANALYZER

6630-01-344-4058 An instrument designed for determination of the carbon dioxide combining power, oxygen, carbon monoxide and nitrogen of whole blood or blood plasma. May also be used for micro determinations of gas other than blood gases, developed from solutions, such as organic nitrogen and carbon, urea, amino nitrogen, iodates, sulphates, sugar, calcium lactic acid, potassium, and analysis of gas mixtures.

Part Alternates: 6630-01-344-4058, 01-344-4058, 6630013444058, 013444058

Instruments and Laboratory Equipment | Chemical Analysis Instruments

Supply Group (FSG) NSN Assign. NIIN
66 01-344-4058

Request a Quote

What Our Customers Say

Compare

NSNs for Compare ( up to 4 ): Add 6630-01-344-4058

Technical Data | NSN 6630-01-344-4058

Characteristic Specifications
METHOD OF ANALYSIS POLAROGRAPHIC AND POTENTIOMETRIC/ION SELECTIVE ELECTRODE
TEST TYPE FOR WHICH DESIGNED BLOOD GAS/ELECTROLYTES; HC03; HCT; BASE EXCESS; TC02 AND 02 SAT
ANALYSIS TEST/MEASUREMENT RANGE PO2, 0 TO 600 MMHG; PH, 6.8 TO 7.8; PCO2, 8 TO 99 MMHG; K, 1.0 TO 9.9 MM/L; NA, 100 TO 200 MM/L; CA, 0.25 TO 4.25 MM/L; HCT, 10 TO 75 PCT; HCO3, 3.0 TO 60MM/L; TCO2, 3.0 TO 60MM/L; BASE EXCESS, -30 TO +30 MM/L; O2 SAT, 30 TO 99 PCT
TEST SAMPLE/SPECIMEN SIZE 0.5 CC
TEST SAMPLE/SPECIMEN TYPE HEPARINIZED WHOLE BLOOD
FEATURES PROVIDED DIGITAL DISPLAY SCREEN AND INTEGRAL PRINTER AND SELF-CALIBRATION
CURRENT TYPE AC
VOLTAGE IN VOLTS 120.0
FREQUENCY RATING 50.0 HERTZ NOMINAL OR 60.0 HERTZ NOMINAL
HEIGHT 8.400 INCHES
WIDTH 9.100 INCHES
DEPTH 18.300 INCHES
UNIT OF ISSUE WEIGHT 27.00 POUNDS
UNIT OF ISSUE CUBE 1.00 CUBIC FEET
SPECIAL FEATURES UTILIZES A DISPOSABLE CARTRIDGE/TEST PACK THAT CONTAINS ALL ELECTRODES, CALIBRATING SOLUTIONS AND GASES PLUS A WASTE BAG FOR PROCESSED BLOOD AND CALIBRATING AND QC SOLUTIONS. THE CARTRIDGE/TEST PACK SHALL ELIMINATE THE NEED FOR GAS TANKS, EXTERNAL SOLUTIONS AND ELECTRODE MAINTENANCE. SHALL BE PRE-PROGRAMED TO AUTOMATICALLY PERFORM TWO-POINT CALLIBRATIONS EVERY HOUR AND ONE-POINT CALIBRATIONS EVERY 90 SECONDS. SHALL AUTOMATICALLY ASPIRATE THE SAMPLE, PROCESS, DISPLAY, AND PRINT THE RESULTS IN LESS THAN TWO MINUTES.
III PART NAME ASSIGNED BY CONTROLLING AGENCY ANALYZER BLOOD GAS 50/60 HZ 120 & 230 V 8.4X9.1X18.3" PORTABLE

Restrictions/Controls & Freight Information | NSN 6630-01-344-4058

Category Code Description
Hazardous Material Indicator CodeNThere is no data in the HMIS and the NSN is in an FSC not generally suspected of containing hazardous materials
Category Code Description
No Freight Information