NSN 6630-01-344-4058 01-344-4058, 013444058
Product Details | BLOOD GAS ANALYZER
6630-01-344-4058 An instrument designed for determination of the carbon dioxide combining power, oxygen, carbon monoxide and nitrogen of whole blood or blood plasma. May also be used for micro determinations of gas other than blood gases, developed from solutions, such as organic nitrogen and carbon, urea, amino nitrogen, iodates, sulphates, sugar, calcium lactic acid, potassium, and analysis of gas mixtures.
Part Alternates: 6630-01-344-4058, 01-344-4058, 6630013444058, 013444058
Instruments and Laboratory Equipment | Chemical Analysis Instruments
Supply Group (FSG) | NSN Assign. | NIIN | Item Name Code (INC) |
---|---|---|---|
66 | 01-344-4058 | 38049 ( ANALYZER, BLOOD GAS ) |
Request a Quote
What Our Customers Say
Compare
NSNs for Compare ( up to 4 ): Add 6630-01-344-4058
Related Products | NSN 6630-01-344-4058
Technical Data | NSN 6630-01-344-4058
Characteristic | Specifications |
---|---|
METHOD OF ANALYSIS | POLAROGRAPHIC AND POTENTIOMETRIC/ION SELECTIVE ELECTRODE |
TEST TYPE FOR WHICH DESIGNED | BLOOD GAS/ELECTROLYTES; HC03; HCT; BASE EXCESS; TC02 AND 02 SAT |
ANALYSIS TEST/MEASUREMENT RANGE | PO2, 0 TO 600 MMHG; PH, 6.8 TO 7.8; PCO2, 8 TO 99 MMHG; K, 1.0 TO 9.9 MM/L; NA, 100 TO 200 MM/L; CA, 0.25 TO 4.25 MM/L; HCT, 10 TO 75 PCT; HCO3, 3.0 TO 60MM/L; TCO2, 3.0 TO 60MM/L; BASE EXCESS, -30 TO +30 MM/L; O2 SAT, 30 TO 99 PCT |
TEST SAMPLE/SPECIMEN SIZE | 0.5 CC |
TEST SAMPLE/SPECIMEN TYPE | HEPARINIZED WHOLE BLOOD |
FEATURES PROVIDED | DIGITAL DISPLAY SCREEN AND INTEGRAL PRINTER AND SELF-CALIBRATION |
CURRENT TYPE | AC |
VOLTAGE IN VOLTS | 120.0 |
FREQUENCY RATING | 50.0 HERTZ NOMINAL OR 60.0 HERTZ NOMINAL |
HEIGHT | 8.400 INCHES |
WIDTH | 9.100 INCHES |
DEPTH | 18.300 INCHES |
UNIT OF ISSUE WEIGHT | 27.00 POUNDS |
UNIT OF ISSUE CUBE | 1.00 CUBIC FEET |
SPECIAL FEATURES | UTILIZES A DISPOSABLE CARTRIDGE/TEST PACK THAT CONTAINS ALL ELECTRODES, CALIBRATING SOLUTIONS AND GASES PLUS A WASTE BAG FOR PROCESSED BLOOD AND CALIBRATING AND QC SOLUTIONS. THE CARTRIDGE/TEST PACK SHALL ELIMINATE THE NEED FOR GAS TANKS, EXTERNAL SOLUTIONS AND ELECTRODE MAINTENANCE. SHALL BE PRE-PROGRAMED TO AUTOMATICALLY PERFORM TWO-POINT CALLIBRATIONS EVERY HOUR AND ONE-POINT CALIBRATIONS EVERY 90 SECONDS. SHALL AUTOMATICALLY ASPIRATE THE SAMPLE, PROCESS, DISPLAY, AND PRINT THE RESULTS IN LESS THAN TWO MINUTES. |
III PART NAME ASSIGNED BY CONTROLLING AGENCY | ANALYZER BLOOD GAS 50/60 HZ 120 & 230 V 8.4X9.1X18.3" PORTABLE |
Restrictions/Controls & Freight Information | NSN 6630-01-344-4058
Category | Code | Description |
---|---|---|
Hazardous Material Indicator Code | N | There is no data in the HMIS and the NSN is in an FSC not generally suspected of containing hazardous materials |
Category | Code | Description |
---|---|---|
No Freight Information |