Product Details | SODIUM-POTASSIUM ANALYZER

6630-01-242-4050 An automated computer controlled instrument employing ion-selective electrode for measuring sodium and potassium in whole blood, serum, plasma, and urine.

Part Alternates: MODEL 1020, MODEL1020, 210201, 6630-01-242-4050, 01-242-4050, 6630012424050, 012424050

Instruments and Laboratory Equipment | Chemical Analysis Instruments

Supply Group (FSG) NSN Assign. NIIN
66 11-OCT-86 01-242-4050

Request a Quote

What Our Customers Say

Compare

NSNs for Compare ( up to 4 ): Add 6630-01-242-4050

Technical Data | NSN 6630-01-242-4050

Characteristic Specifications
METHOD OF ANALYSIS POTENTIOMETRIC/ION SELECTIVE ELECTRODE
ELECTRODE TYPE POTASSIUM/K+ AND REFERENCE AND SODIUM/NA+
ANALYSIS TEST/MEASUREMENT RANGE SODIUM 100-170 MMOL/L IN BLOOD AND 14-1400 MMOL/L IN UNDILUTED URINE; POTASSIUM 2-7 MMOL/L IN BLOOD AND 0.4-400 MMOL/L IN UNDILUTED URINE
TEST SAMPLE/SPECIMEN SIZE 100 MICROLITERS OF BLOOD OR 400 MICROLITERS OF PREDILUTED URINE
TEST SAMPLE/SPECIMEN TYPE WHOLE BLOOD,SERUM,PLASMA,AND URINE
TEST SAMPLE/SPECIMEN DELIVERY DEVICE NEEDLE/PROBE, ANALYZER ASPIRATION SYSTEM
FEATURES PROVIDED DIGITAL DISPLAY SCREEN AND INTEGRAL PRINTER AND SELF-CALIBRATION
CURRENT TYPE AC
VOLTAGE IN VOLTS 120.0 OR 230.0
FREQUENCY RATING 50.0 HERTZ NOMINAL OR 60.0 HERTZ NOMINAL
HEIGHT 360.0 MILLIMETERS
WIDTH 260.0 MILLIMETERS
DEPTH 230.0 MILLIMETERS
ACCESSORY COMPONENTS AND QUANTITY POTASSIUM ELECTRODE,1; SODIUM ELECTRODE,1; REFERENCE ELECTRODE,1; BOTTLE OF STANDARD A,2; BOTTLE OF STANDARD B,1; WASTE BOTTLE,1; BOTTLE OF REFERENCE FILLING SOLUTION,1; BOTTLE OF TEST DYE SOLUTION,1; REFERENCE MEMBRANE ASSEMBLY,2; FLUID PATH MODULE,1; BOTTLE ELECTRODE RINSE SOLUTION,1; SLIDE VALVE,1; SAMPLE PROBE ASSEMLBY,1
UNIT OF ISSUE WEIGHT 29.00 POUNDS
UNIT OF ISSUE CUBE 2.820 CUBIC FEET
SPECIAL FEATURES IF THE ANALYZER DOES NOT CONTAIN AN INTEGRAL PRINTER,IT SHALL BE CAPABLE OF INTERFACING W/A COMMERCIALLY AVAILABLE EXTERNAL PRINTER
III PART NAME ASSIGNED BY CONTROLLING AGENCY ANALYZER POTASSIUM AND SODIUM ION SELECTIVE:

Restrictions/Controls & Freight Information | NSN 6630-01-242-4050

Category Code Description
Shelf-Life Code:624 months
Hazardous Material Indicator CodeNThere is no data in the HMIS and the NSN is in an FSC not generally suspected of containing hazardous materials
Demilitarization Code:ANon-Munitions List Item/ Non-Strategic List Item - Demilitarization not required.
Controlled Inventory Item Code:UUNCLASSIFIED
Precious Metals Indicator Code:AItem does not contain precious metal
Criticality Code:X The item does not have a nuclear hardened feature or any other critical feature such as tolerance, fit restriction or application.
Automatic Data Processing Equipment:0Represents items with no ADP components. NOTE: Codes 1 through 6 are only to be used when the item is Automatic Data Processing Equipment (ADPE) in its entirety and is limited to the type meeting only one of the definitions for codes 1 through 6. (See code 9)
Category Code Description
No Freight Information